Journal article icon

Journal article

Addressing a systematic vibration artifact in diffusion-weighted MRI.

Abstract:

We have identified and studied a pronounced artifact in diffusion-weighted MRI on a clinical system. The artifact results from vibrations of the patient table due to low-frequency mechanical resonances of the system which are stimulated by the low-frequency gradient switching associated with the diffusion-weighting. The artifact manifests as localized signal-loss in images acquired with partial Fourier coverage when there is a strong component of the diffusion-gradient vector in the left-righ...

Expand abstract
Publication status:
Published

Actions


Access Document


Publisher copy:
10.1002/hbm.20856

Authors


Gallichan, D More by this author
Bartsch, A More by this author
More by this author
Institution:
University of Oxford
Department:
Oxford, MSD, Clinical Neuroscience, Experimental Psychology
More by this author
Institution:
University of Oxford
Department:
Oxford, MSD, RDM, Cardiovascular Medicine, BHF Centre of Research Excellence
Expand authors...
Journal:
Human brain mapping
Volume:
31
Issue:
2
Pages:
193-202
Publication date:
2010-02-05