Development and validation of a liquid chromatographic-tandem mass spectrometric method for determination of piperaquine in plasma stable isotope labeled internal standard does not always compensate for matrix effects.

Abstract:

A bioanalytical method for the analysis of piperaquine in human plasma using off-line solid-phase extraction and liquid chromatography coupled to positive tandem mass spectroscopy has been developed and validated. It was found that a mobile phase with high pH (i.e. 10) led to better sensitivity than mobile phase combinations with low pH (i.e. 2.5-4.5) despite the use of positive electrospray and a basic analyte. The method was validated according to published FDA guidelines and showed excelle...

Expand abstract
Publication status:
Published

Actions


Access Document


Authors


More by this author
Institution:
University of Oxford
Division:
MSD
Department:
NDM
Role:
Author
More by this author
Institution:
University of Oxford
Division:
MSD
Department:
NDM
Role:
Author
Journal:
Journal of chromatography. B, Analytical technologies in the biomedical and life sciences More from this journal
Volume:
862
Issue:
1-2
Pages:
227-236
Publication date:
2008-02-01
DOI:
EISSN:
1873-376X
ISSN:
1570-0232
Language:
English
Keywords:
Pubs id:
pubs:12435
UUID:
uuid:21dee465-af26-4c97-9543-272629d471c2
Local pid:
pubs:12435
Source identifiers:
12435
Deposit date:
2012-12-19

Terms of use


Views and Downloads






If you are the owner of this record, you can report an update to it here: Report update to this record

TO TOP