Thesis
Accurate and precise 3D T1 mapping of the human liver at 3T using widely available magnetic resonance imaging methods
- Abstract:
-
Better diagnostic and treatment methods are needed to counteract the steep increase of liver diseases. A promising non-invasive biomarker for the diagnosis of liver diseases is the longitudinal relaxation time (T1) of liver tissue, measured through quantitative magnetic resonance imaging. T1 correlates with liver fibrosis and inflammation. However, non-contrast T1 mapping is currently restricted to a single slice cross sectioning the liver and uses specialist cardiac imaging sequences that are often unavailable. As such, the main questions addressed in this thesis were: Is it possible to map T1 across the whole liver using only widely available pulse sequences? If so, can one obtain an unbiased estimate of liver T1 with sufficient precision for the diagnosis of liver diseases?
To answer these questions, a set of T1-weighted sequences capable of achieving whole liver coverage was adopted. The large covariance between T1 and the transmitted magnetic field (B1+) highlighted the need for an accurate B1+ map. Slice profile effects and several off-resonance based B1+ confounders were identified and corrected. Uncorrected, the median B1+ error in a phantom was -5.2%, producing a median T1 error of 10.5%. After applying all corrections, the median T1 error decreased to -0.3% with an interquartile range of [-2.5 2.1]%.
A theoretical framework was developed to calculate the T1 variance, including the noise propagated from the B1+ map into the T1 map. This framework was validated in vivo and used to find the optimal flip angles that resulted in the highest precision across a range of T1s and B1+ factors observed in the liver at 3 T.
Using scanner-agnostic sequences together with corrections for the confounding factors, a variable flip angle 3D T1 mapping method was developed requiring 7 breath-holds. A study across 10 healthy volunteers at 3 T validated the method’s accuracy, scan/rescan repeatability and precision. The 3D T1 maps yielded mean T1 values that, on average, were 5% lower than alternative single-slice T1 maps, with a repeatability coefficient of 49 ms and a median T1 standard deviation of 31 ms, showing potential to diagnose liver fibrosis.
Actions
Access Document
- Files:
-
-
(Preview, Dissemination version, pdf, 28.4MB, Terms of use)
-
Authors
Contributors
- Institution:
- University of Oxford
- Division:
- MSD
- Department:
- RDM
- Sub department:
- RDM Cardiovascular Medicine
- Role:
- Supervisor
- ORCID:
- 0000-0002-6945-5201
- Institution:
- University of Oxford
- Division:
- MSD
- Department:
- RDM
- Sub department:
- RDM Cardiovascular Medicine
- Research group:
- Tyler Group
- Oxford college:
- Somerville College
- Role:
- Supervisor
- ORCID:
- 0000-0002-0780-8905
- Institution:
- Perspectum and University of Oxford
- Division:
- MSD
- Department:
- RDM
- Sub department:
- RDM Cardiovascular Medicine
- Role:
- Supervisor
- Grant:
- EP/R512333/1
- Grant:
- EP/R512333/1
- Programme:
- Oxford-Nottingham Biomedical Imaging CDT
- DOI:
- Type of award:
- DPhil
- Level of award:
- Doctoral
- Awarding institution:
- University of Oxford
- Language:
-
English
- Keywords:
- Subjects:
- Deposit date:
-
2022-07-21
- ARK identifier:
Terms of use
- Copyright holder:
- De Matos Gomes Belsley, G
- Copyright date:
- 2022
If you are the owner of this record, you can report an update to it here: Report update to this record