Thesis
Optimising hyperpolarised carbon-13 MRI for metabolic imaging: a combined approach of hardware innovation, accelerated acquisitions and denoised spectra
- Abstract:
-
Hyperpolarised (HP) carbon-13 MRI offers a unique, non-invasive window into in vivo metabolism by tracking the enzymatic conversion of [1-13C]pyruvate into downstream metabolites such as lactate, alanine, and bicarbonate. However, its clinical translation is constrained by the rapid, irreversible decay of the enhanced signal, the low sensitivity of the 13C nucleus, and the narrow acquisition window within which spatial, spectral, and temporal information must be recovered. This thesis addresses three bottlenecks in the HP 13C imaging chain: receive sensitivity, acquisition speed, and quantification robustness.
First, I developed a four-element multi-layer receive array for cardiac HP 13C MRI at 3 T. Through electromagnetic simulations, bench-top S-parameter measurements, and in-bore Bloch-Siegert B + 1 mapping, I identified an 11 cm centre-to-centre spacing that prioritised field homogeneity across a heart-sized field-of-view. The array achieved a mean SNR of 14.68 ± 4.79, outperforming a commercial rigid chest array (12.96 ± 3.75), and delivered more than twice the mean receive sensitivity at depth.
econd, I implemented SENSitivity Encoding (SENSE) to accelerate 2D chemical shift imaging (CSI) fourfold, reducing acquisition time from 36 s to 9 s per time point. Using a dual-tuned 23Na/13C flexible receive array, I used naturally abundant sodium sensitivity maps as a surrogate for 13C coil sensitivities, validated in phantom and porcine kidney studies.
Third, I developed a dynamic spectral fitting algorithm that leverages high-SNR summed spectra as prior knowledge to constrain the fitting of individual low-SNR time points. Validated across rat, mouse, and human datasets, the dynamic fit reduced the mean Cramér-Rao Lower Bound by up to 78% and kinetic model RMSE by up to 76%, enabling a 79% reduction in the sample size required to detect metabolic differences in a mouse model of obesity.
Actions
Access Document
- Files:
-
-
(Preview, Dissemination version, pdf, 55.2MB, Terms of use)
-
Authors
Contributors
+ Tyler, D
- Institution:
- University of Oxford
- Division:
- MSD
- Department:
- Physiology Anatomy and Genetics
- Oxford college:
- Somerville College
- Role:
- Supervisor
- ORCID:
- 0000-0002-0780-8905
+ Grist, J
- Institution:
- University of Oxford
- Division:
- MSD
- Department:
- Radcliffe Department of Medicine
- Role:
- Supervisor
- ORCID:
- 0000-0001-7223-4031
+
University of Oxford
More from this funder
- Grant:
- BRT00030
- Programme:
- MRC Enterprise (Industrial CASE) studentship, Clarendon Scholarship, Department of Physiology, Anatomy, and Genetics
- DOI:
- Type of award:
- DPhil
- Level of award:
- Doctoral
- Awarding institution:
- University of Oxford
- Language:
-
English
- Deposit date:
-
2026-08-01
- ARK identifier:
Terms of use
- Copyright holder:
- Ayaka Shinozaki
- Copyright date:
- 2026
If you are the owner of this record, you can report an update to it here: Report update to this record