Thesis icon

Thesis

Vital sign monitoring and data fusion in haemodialysis

Abstract:

Intra-dialytic hypotension (IDH) is the most common complication in haemodialysis (HD) treatment and has been linked with increased mortality in HD patients. Despite various approaches towards understanding the underlying physiological mechanisms giving rise to IDH, the causes of IDH are poorly understood. Heart Rate Variability (HRV) has previously been suggested as a predictive measure of IDH. In contrast to conventional spectral HRV measures in which the frequency bands are defined by fixed limits, a new spectral measure of HRV is introduced in which the breathing rate is used to identify and measure the physiologically-relevant peaks of the frequency spectrum. The ratio of peaks leading up to the IDH event was assessed as a possible measure for IDH prediction. Changes in the proposed measure correlate well with the magnitude of abrupt changes in blood pressure in patients with autonomic dysfunction, but there is no such correlation in patients without autonomic dysfunction.

At present, routine clinical vital sign monitoring beyond simple weight and blood pressure measurements at the start and end of each session has not established itself in clinical practice. To investigate the benefits of continuous vital sign monitoring in HD patients with regard to detecting and predicting IDH, different population-based and patient-specific models of normality were devised and tested on data from an observational study at the Oxford Renal Unit in which vital signs were recorded during HD sessions. Patient-specific models of normality performed better in distinguishing between IDH and non-IDH data, primarily due to the wide range of vital sign data included as part of the training data in the population-based models. Further, a patient-specific data fusion model was constructed using Parzen windows to estimate a probability density function from the training data consisting of vital signs from IDH-free sessions. Although the model was constructed using four vital sign inputs, novelty detection was found to be primarily driven by blood pressure decreases.

Actions

Access Document

Files:

Authors

More by this author
Institution:
University of Oxford
Division:
MPLS
Department:
Engineering Science
Research group:
Biomedical Signal Processing and m-Health
Oxford college:
Queen's College
Role:
Author

Contributors

Division:
MPLS
Department:
Engineering Science
Role:
Supervisor


Engineering and Physical Sciences Research Council More from this funder
Funding agency for:
Borhani, Y


Publication date:
2013
DOI:
Type of award:
DPhil
Level of award:
Doctoral
Awarding institution:
University of Oxford


Language:
English
Keywords:
Subjects:
UUID:
uuid:8c36b904-0558-4161-b756-1a2dd910830f
Local pid:
ora:8845
Deposit date:
2014-07-30
ARK identifier:

Terms of use


Views and Downloads

Views and downloads will return soon






If you are the owner of this record, you can report an update to it here: Report update to this record