Non-contact detection of cessations of breathing in pre-term infants
- Abstract:
-
Current technologies for detecting Cessation Of Breathing (COBE) episodes in pre-term infants, such as Impedance Pneumography (IP) and electrocardiography (ECG), require direct skin contact through adhesive electrodes. These contact-based methods can cause discomfort and skin irritation in pre-term infants with fragile skin and are prone to motion artefacts, resulting in false alarms and alarm fatigue that can compromise infant safety. Video-based physiological monitoring offers a non-contact alternative for continuous respiratory assessment without disrupting routine care. However, these methods have largely been developed and validated in adults, and their translation to pre-term infants, who present distinct physiological and care-related challenges, remains underexplored.
This thesis proposes and evaluates a framework for non-contact detection of COBE events in pre-term infants. I developed and validated the methodology using data from a clinical study of 30 pre-term infants hospitalised in the high dependency area of the Neonatal Intensive Care Unit (NICU) at the John Radcliffe Hospital, Oxford. The dataset comprised 90 recording sessions of RGB video recordings and reference physiological signals collected under routine clinical conditions, totalling over 430 hours.
I extracted respiratory information from video recordings during periods characterised by motion and physiological instability surrounding COBE episodes. I developed camera-based respiratory rate estimation methods with motion-based filter switching to ensure reliable performance in the NICU. I then trained deep learning models to detect COBE episodes using contact-based physiological signals, video-derived signals, and their combinations. Models were developed using five-fold cross-validation on 83.3% of the dataset, with 16.7% reserved as an independent held-out test set. Among contact-only models, ConvNeXt achieved the highest performance on combined IP and photoplethysmography (PPG) signals, with a test accuracy of 88.7%. Hybrid models integrating IP with video-derived pixel intensity achieved the highest performance (90.6%), outperforming all unimodal approaches. Camera-only models achieved a test accuracy of 76.9%, outperforming models trained on ECG or PPG alone.
These results demonstrate the feasibility of non-contact COBE detection in pre-term infants and show that integrating video-derived features with traditional physiological signals improves detection performance. The proposed hybrid framework enhances reliability without replacing existing monitoring infrastructure and supports improved respiratory monitoring in high-resource NICU settings. Adapting this approach to widely available devices, such as smartphone cameras, could further enable scalable monitoring in low-resource environments where conventional monitoring equipment may not be available.
Actions
Access Document
- Files:
-
-
(Preview, Dissemination version, pdf, 36.1MB, Terms of use)
-
Authors
Contributors
+ Tarassenko, L
- Institution:
- University of Oxford
- Division:
- MPLS
- Department:
- Engineering Science
- Role:
- Supervisor
+ Villarroel Montoya, M
- Institution:
- University of Oxford
- Division:
- MPLS
- Department:
- Engineering Science
- Role:
- Supervisor
- DOI:
- Type of award:
- DPhil
- Level of award:
- Doctoral
- Awarding institution:
- University of Oxford
- Language:
-
English
- Keywords:
- Subjects:
- Deposit date:
-
2026-08-06
- ARK identifier:
Terms of use
- Copyright holder:
- Dineo Serame
- Copyright date:
- 2026
If you are the owner of this record, you can report an update to it here: Report update to this record