Thesis icon

Thesis

Parallel transmission technologies for imaging the brain and body with 7 tesla MRI

Abstract:

The push to higher-and-higher static magnetic field strengths for magnetic resonance imaging systems leads to the radiofrequency fields, that are used for image acquisition, becoming less-and-less uniform. This results in undesirable changes in image contrast which are not due to anatomy. It is, however, possible to mitigate for this radiofrequency field non-uniformity using technology known as parallel transmission. Although, in order to provide a robust subject-specific improvement the non-uniform radiofrequency field must first be mapped in a process known as “B1+ mapping”.

To begin, I will go through some essential magnetic resonance physics and cover some of the key concepts used in this thesis. I will then introduce ultra-high field magnetic resonance imaging and discuss why it is of interest and some of the challenges it currently faces.

The first research chapter explores a new way of mapping this non-uniform radiofrequency field using a novel adaption to a previously used pulse sequence. This method, referred to as "Sandwich", was successful in producing absolute B1+ maps across the entire thorax at 7 T and in particular in the heart where typically many methods fail due to blood flow. Using extended phase graphs and Bloch simulations I investigated how this new method compares to those currently in use. I also investigated whether coil-cycling the acquisition of relative maps was beneficial and I found that coil-cycling is especially important when measuring 3D relative magnitude maps.

Mapping many individual transmit channels is time-consuming, hence a method to accelerate the acquisition of B1+ maps for the entire transmit array over a large volume was also explored. I did this by combining two methods known as "TxLR" and "B1TIAMO" with the Sandwich method mentioned previously. I showed that accurate B1+ maps can be reconstructed from low-resolution and sparse k-space data. I demonstrated this using pulse sequence capable of acquiring 3D single-channel maps for an 8-channel transmit array in only 14 seconds in the brain or 26 seconds in the body.

Finally, subject motion has the ability to degrade the performance of parallel transmission technology. Knowledge of the subject’s motion can help inform both the acquisition and reconstruction for motion correction. Motion detection can be achieved through various methods, but radiofrequency sensor-based techniques like parallel transmit scattering or pilot tone are of particular interest, primarily due to their low hardware requirements. Hence, the final research chapter explores these two competing radiofrequency sensor-based methods for motion detection. Both parallel transmit scattering and pilot tone showed good sensitivity to head motion and each method was able to accurately predict rigid body head motion in all six degrees of freedom using a linear regression model.

Actions

Access Document

Files:

Authors

More by this author
Institution:
University of Oxford
Division:
MSD
Department:
Clinical Neurosciences
Sub department:
Clinical Neurosciences
Research group:
WIN Physics
Oxford college:
Lincoln College
Role:
Author
ORCID:
https://orcid.org/0000-0001-6482-7515

Contributors

Institution:
University of Oxford
Division:
MSD
Department:
Clinical Neurosciences
Sub department:
Clinical Neurosciences
Research group:
WIN Physics
Role:
Supervisor
ORCID:
https://orcid.org/0000-0002-9289-5619
Institution:
University of Oxford
Division:
MSD
Department:
RDM
Sub department:
RDM Cardiovascular Medicine
Role:
Supervisor
ORCID:
https://orcid.org/0000-0003-2567-3642
Role:
Supervisor


UK Research and Innovation More from this funder
Funding agency for:
Kent, JL
Grant:
2432150
Programme:
EPSRC iCASE


DOI:
Type of award:
DPhil
Level of award:
Doctoral
Awarding institution:
University of Oxford


Terms of use


Views and Downloads

Views and downloads will return soon






If you are the owner of this record, you can report an update to it here: Report update to this record

TO TOP